Abbott Laboratories automatic sensing control algorithms
Pictorial diagram showing how SenseAbility, the <t>automatic</t> <t>sensing</t> <t>control</t> <t>algorithm,</t> worked in our patient with the left bundle branch pacing lead connected to the atrial port. Following the paced R wave and the blanking period, the post-paced sensitivity threshold started at 0.8 mV (automatically set by device). As the decay delay is nominally 0 ms, this sensitivity threshold immediately starts to decrease until it reaches the maximum sensitivity programmed (0.2 mV in this case) over the programmed decay duration of 220 ms. Here, T wave oversensing is demonstrated. ECG = electrocardiogram.
Automatic Sensing Control Algorithms, supplied by Abbott Laboratories, used in various techniques. Bioz Stars score: 86/100, based on 1 PubMed citations. ZERO BIAS - scores, article reviews, protocol conditions and more
https://www.bioz.com/product/automatic+sensing+control+algorithms/pmc12666912-0-0-5?v=Abbott+Laboratories
Average 86 stars, based on 1 article reviews
automatic sensing control algorithms - by Bioz Stars, 2026-07
86/100 stars

Images

1) Product Images from "Terrible T wave oversensing—programming pitfalls and optimization of the dual chamber implantable cardioverter-defibrillator with left bundle branch pacing"

Article Title: Terrible T wave oversensing—programming pitfalls and optimization of the dual chamber implantable cardioverter-defibrillator with left bundle branch pacing

Journal: HeartRhythm Case Reports

doi: 10.1016/j.hrcr.2025.08.019

Pictorial diagram showing how SenseAbility, the automatic sensing control algorithm, worked in our patient with the left bundle branch pacing lead connected to the atrial port. Following the paced R wave and the blanking period, the post-paced sensitivity threshold started at 0.8 mV (automatically set by device). As the decay delay is nominally 0 ms, this sensitivity threshold immediately starts to decrease until it reaches the maximum sensitivity programmed (0.2 mV in this case) over the programmed decay duration of 220 ms. Here, T wave oversensing is demonstrated. ECG = electrocardiogram.
Figure Legend Snippet: Pictorial diagram showing how SenseAbility, the automatic sensing control algorithm, worked in our patient with the left bundle branch pacing lead connected to the atrial port. Following the paced R wave and the blanking period, the post-paced sensitivity threshold started at 0.8 mV (automatically set by device). As the decay delay is nominally 0 ms, this sensitivity threshold immediately starts to decrease until it reaches the maximum sensitivity programmed (0.2 mV in this case) over the programmed decay duration of 220 ms. Here, T wave oversensing is demonstrated. ECG = electrocardiogram.

Techniques Used: Control



Similar Products

86
Abbott Laboratories automatic sensing control algorithms
Pictorial diagram showing how SenseAbility, the <t>automatic</t> <t>sensing</t> <t>control</t> <t>algorithm,</t> worked in our patient with the left bundle branch pacing lead connected to the atrial port. Following the paced R wave and the blanking period, the post-paced sensitivity threshold started at 0.8 mV (automatically set by device). As the decay delay is nominally 0 ms, this sensitivity threshold immediately starts to decrease until it reaches the maximum sensitivity programmed (0.2 mV in this case) over the programmed decay duration of 220 ms. Here, T wave oversensing is demonstrated. ECG = electrocardiogram.
Automatic Sensing Control Algorithms, supplied by Abbott Laboratories, used in various techniques. Bioz Stars score: 86/100, based on 1 PubMed citations. ZERO BIAS - scores, article reviews, protocol conditions and more
https://www.bioz.com/product/automatic+sensing+control+algorithms/pmc12666912-0-0-5?v=Abbott+Laboratories
Average 86 stars, based on 1 article reviews
automatic sensing control algorithms - by Bioz Stars, 2026-07
86/100 stars
  Buy from Supplier

Image Search Results


Pictorial diagram showing how SenseAbility, the automatic sensing control algorithm, worked in our patient with the left bundle branch pacing lead connected to the atrial port. Following the paced R wave and the blanking period, the post-paced sensitivity threshold started at 0.8 mV (automatically set by device). As the decay delay is nominally 0 ms, this sensitivity threshold immediately starts to decrease until it reaches the maximum sensitivity programmed (0.2 mV in this case) over the programmed decay duration of 220 ms. Here, T wave oversensing is demonstrated. ECG = electrocardiogram.

Journal: HeartRhythm Case Reports

Article Title: Terrible T wave oversensing—programming pitfalls and optimization of the dual chamber implantable cardioverter-defibrillator with left bundle branch pacing

doi: 10.1016/j.hrcr.2025.08.019

Figure Lengend Snippet: Pictorial diagram showing how SenseAbility, the automatic sensing control algorithm, worked in our patient with the left bundle branch pacing lead connected to the atrial port. Following the paced R wave and the blanking period, the post-paced sensitivity threshold started at 0.8 mV (automatically set by device). As the decay delay is nominally 0 ms, this sensitivity threshold immediately starts to decrease until it reaches the maximum sensitivity programmed (0.2 mV in this case) over the programmed decay duration of 220 ms. Here, T wave oversensing is demonstrated. ECG = electrocardiogram.

Article Snippet: Automatic sensing control algorithms , Abbott , Medtronic , Biotronik , Boston Scientific.

Techniques: Control